Siemens · Iconos MD

Siemens Iconos MD: stretched image with pincushion distortion in full format

What happened

Reported in the case: in full format mode the image was stretched — a 22 cm reference object displayed as about 24 cm — with strong pincushion distortion, while zoom 1 and zoom 2 looked normal.

What was checked

Cause

The engineers found the full-format electronic optics of the image intensifier out of adjustment: the accelerating voltages and anode focus for that format were off, so the image came out larger with pincushion distortion. Zooms 1 and 2 use their own optics settings, which is why they stayed correct.

What was done

The engineers tuned the electronic optics in full format mode: with the service potentiometers they reset the accelerating voltages and the anode focus — essentially the same alignment as on a CRT tube, done per the manufacturer's service documentation by authorised personnel. Small residual nonlinear distortion near the edges remained but was accepted as within tolerance.

Safety

Geometry checks mean the X-ray beam is on with a test object in the field: keep screening time and dose as low as possible, keep hands and staff out of the primary beam, and confirm geometry and image quality on a phantom before returning the unit to clinical use.

Lesson: When a size error appears in only one format of an II-based fluoro system, suspect that format's electronic optics alignment and verify geometry per format before swapping hardware or software.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.