Siemens Iconos MD: stretched image with pincushion distortion in full format
What happened
Reported in the case: in full format mode the image was stretched — a 22 cm reference object displayed as about 24 cm — with strong pincushion distortion, while zoom 1 and zoom 2 looked normal.
What was checked
- Verified image geometry with a known reference object (reported as a 22 cm screwdriver) in every format: only full format was out, zooms 1 and 2 were correct.
- Replaced the image intensifier complete with its camera unit — the full-format fault remained.
- Updated the Fluorospot software, which did not change the behaviour.
- Discussed where the format magnification actually happens (in the II/FLC chain or in the XCU) and concluded the full-format optics sit in the image intensifier chain.
Cause
The engineers found the full-format electronic optics of the image intensifier out of adjustment: the accelerating voltages and anode focus for that format were off, so the image came out larger with pincushion distortion. Zooms 1 and 2 use their own optics settings, which is why they stayed correct.
What was done
The engineers tuned the electronic optics in full format mode: with the service potentiometers they reset the accelerating voltages and the anode focus — essentially the same alignment as on a CRT tube, done per the manufacturer's service documentation by authorised personnel. Small residual nonlinear distortion near the edges remained but was accepted as within tolerance.
Safety
Geometry checks mean the X-ray beam is on with a test object in the field: keep screening time and dose as low as possible, keep hands and staff out of the primary beam, and confirm geometry and image quality on a phantom before returning the unit to clinical use.
Lesson: When a size error appears in only one format of an II-based fluoro system, suspect that format's electronic optics alignment and verify geometry per format before swapping hardware or software.