Siemens Iconos R100: errors when entering tomography mode
What happened
As reported in the case, every time the unit was switched into tomography mode the console raised errors 223/240 and 422/240. The first working theory was that the console did not see the cassette, but the cassette was in place and the errors still came up.
What was checked
- Checked that the cassette was correctly inserted and seated — in this case it was, so the cassette was ruled out.
- Noticed that the errors appeared only on entering tomography mode, with other modes working as reported — this pointed to the mode itself being locked rather than to a fault in the imaging chain (tube, detector, generator).
Cause
The tomography function was locked on the console — an activation/permission state of the option, not a hardware failure. While the lock was in place, the reported error pair came up on every tomography request.
What was done
The tomography lock was removed as a procedure from the manufacturer's service documentation, done by authorised personnel. After that, as stated in the case, the unit entered tomography mode without the reported errors.
Safety
Before returning the unit to clinical use, confirm that exposure starts and terminates normally from every control, that interlocks and warning signals work, and keep the room clear during any test exposure.
Lesson: Mode-selective error codes can simply mean the mode is locked or not configured, so check cassette seating and the console's lock/configuration state against the documentation before suspecting a hardware fault.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.