Siemens Iconos R100: stand tilt angle sensor reading out of range
What happened
During normal operation the unit reported error 252-32, decoded in the case as the stand tilt potentiometer M2.R5 reading outside its allowed window — the diagnostic value, as reported, sat at the edge of the range (below 3 or above 252 counts). The practical split noted in the case: a sporadic error points to contact or coupling trouble, a constant one to a hard fault in the sensor chain or on the A4.D41 board.
What was checked
- Read the live tilt-angle value in the unit's diagnostics (access per the manufacturer's service documentation) and watch whether it is pinned at one end of the range, jumps, or follows the stand smoothly.
- Check potentiometer M2.R5 and its mechanical engagement with the stand — the sensor must actually rotate with the movement.
- Inspect the cables and connectors on board A4.D41 for bent pins, oxidation or loose contact, and measure the 5 V feed to the potentiometer reported in the case.
- If the error is constant and the wiring and supply check out, suspect the A4.D41 board itself.
Cause
Not confirmed in the case. Most probable: intermittent contact somewhere in the measurement chain — a worn potentiometer track or loose coupling, a faulty connector or cable to A4.D41, or a fault on the A4.D41 board itself — driving the reading to an end-of-scale value.
The cause is probable and was not fully confirmed.
What was done
Approach reported in the case: for a sporadic error, monitor the live value while moving the stand and re-seat or repair the connectors and the potentiometer coupling; for a constant error, verify M2.R5, its wiring, the connectors at A4.D41 and the 5 V supply, then replace the potentiometer or the A4.D41 board as indicated. Replacement and any recalibration are a procedure from the manufacturer's service documentation, done by authorised personnel.
Safety
Before working on the stand, de-energise the unit per the service documentation and mechanically secure the stand, because the tilted, counterweighted head can drift or swing under gravity once power is removed, and verify the tilt indication over the full travel before returning the unit to use.
Lesson: When an analogue angle sensor reports a rail-edge or jumpy value, prove the supply, wiring and mechanical coupling first and treat the board as the last suspect.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.