Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Siemens Impact (магнит OR42)
Siemens · Impact (магнит OR42)

Siemens Impact (OR42 magnet): cryostat shield temperature check needs a 100 µA source and a missing pinout

What happened

The engineers had to measure the cryostat thermal shield temperatures. The documentation names the sensors and the 100 µA excitation current they need, but the connector pin assignment was not written out, so it was unclear how to connect the meter.

What was checked

Cause

No equipment fault was reported. The gap was practical: the manual describes the shield temperature sensors and the 100 µA current they require, but does not give the connector pinout (all as reported in the case).

The cause is probable and was not fully confirmed.

What was done

As reported, the reliable route is to measure with a calibrated meter of the type the manufacturer recommends, using the pin assignment for the 80 K and 20 K shields and the neck sensors; before connecting, verify the pinout against the manufacturer's service documentation. If such a meter is not available, the engineers' fallback was an improvised 100 µA current source from three 9 V batteries and a resistor chain of about 270 kΩ with a trimmer; the 1.5 V variant was dropped as too inaccurate. This is one reported case, so the outcome is probable rather than proven on other systems.

Safety

An MRI magnet holds its strong static field at all times, even powered down: use non-magnetic tools, keep any improvised battery pack and leads firmly secured, and inspect lead insulation before connecting anything to the cryostat sensor connector.

Lesson: When a routine measurement detail like a sensor pinout is missing from the manual, first take the pinout and excitation source from the manufacturer's service documentation and a calibrated meter, and trust an improvised source only if it genuinely delivers the stated current.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes