Siemens Luminos RF: false HV-tank overpressure alarm
What happened
During operation the system flagged 'Overpressure in the HV-tank' (reported in the case as ERR 40484/80). After the repair a separate error (reported in the case as ERR 40513/80) appeared and was not resolved within this case.
What was checked
- Continuity check of the HV-tank pressure switch inside the generator: no contact through the switch.
- Inspection of the switch solder joint: the joint was poor, causing the open circuit.
Cause
The engineers found a poor solder joint on the HV-tank pressure switch inside the generator. The open switch circuit was read by the generator as overpressure in the HV-tank; the alarm was false, and no actual pressure problem was reported in the case.
What was done
The engineers restored the solder joint on the pressure switch, confirmed continuity, and the system returned to normal operation — reported as solved. The later error reported as ERR 40513/80 is a separate issue that this repair did not cover.
Safety
Before opening the generator, isolate the equipment from the mains and let the HV generator and its capacitors discharge following the manufacturer's procedure — the HV tank can hold a lethal charge long after power-off.
Lesson: An HV-tank overpressure alarm is not always real pressure — verify the pressure switch circuit, including wiring, connectors and solder joints, before condemning the tank.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.