Siemens · Magnetom Aero 1.5T

Siemens Magnetom Aero 1.5T: knee coil reports no resonance, scans abort

What happened

Reported in the case: when a knee study was started, the system reported no resonance from the knee coil and the scan stopped. The behaviour depended on movement — wiggling the cable or the adapter could trigger it.

What was checked

Cause

The engineers found a poor ground contact inside the coil adapter: the joint was breaking intermittently under vibration, so the system stopped seeing a properly resonating coil and aborted the study.

What was done

The adapter was disassembled, the ground contact was cleaned and the joint was re-soldered. After reassembly the fault no longer reproduced, and the study ran normally — reported as solved in this case.

Safety

MRI safety: the static magnetic field is always on — never take ferromagnetic tools or loose metal parts into the magnet room, and service the adapter outside the exam room where possible.

Lesson: An intermittent no-resonance or coil-loss message can come from a simple mechanical contact fault in the adapter, so reproduce it on a phantom by moving cables and connectors before blaming the coil or the system.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.