Siemens Magnetom Avanto: exam database cleanup looks like a hang
What happened
Exam database cleanup on the data workstation ran for about 35–45 minutes (as reported in the case), did not reboot the system on its own and looked like a hang; an early reboot left the database only partly cleaned.
What was checked
- The engineers timed a full cleanup run to establish its normal duration instead of assuming a hang from the start.
- They compared a plain full cleanup with the approach of deleting the studies first and only then starting the cleanup utility.
Cause
In this case there was no fault as such: on this system the cleanup simply runs long and does not restart the workstation by itself, so a reboot in the first minutes interrupted the job and left the database uncleaned.
What was done
Let the cleanup finish before any restart: in this case the engineers started it remotely on a timer for about 30 minutes and only then rebooted the workstation manually. An alternative that also worked here: delete the studies first, then run the AllDBRemove cleanup — as reported, this shortened the job from roughly 35 to 15 minutes. The database was confirmed clean before the system went back to use.
Safety
Run the cleanup only in scheduled downtime with no scan in progress, and confirm the exam database is intact before returning the system to clinical use.
Lesson: A long-running maintenance process is not automatically a hang — time its normal duration first and never interrupt it with a reboot halfway through.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.