Siemens Magnetom: BTB passes with dummy plug but no link to the coil
What happened
The BTB interface module passed its check with a dummy plug in place, but when a coil was connected (reported in the case as the body coil) there was no link at all.
What was checked
- Run the BTB check twice in a row: once with the dummy plug, once with the real coil connected. A pass with the plug and a fail with the coil points to the interface module, not to the coil itself.
- Inspect the board solder joints under magnification; the engineers found cracked legs on capacitors in this case.
- checks_ru
Cause
The engineers found fatigued solder joints: cracks on the capacitor leads broke contact only under the real load of a connected coil, which is why the dummy-plug test still passed.
What was done
The board was recovered by reflowing all the component legs on the affected section. The fault was solved. As reported in the case, colleagues mentioned four similar failures where a leak went together with cracked leads, so checking this module type for traces of moisture is a reasonable habit.
Safety
Capacitors can hold a charge after power-off, so discharge and verify before touching the board, and complete the manufacturer's electrical safety checks (protective earth, leakage current) before the coil path is returned to patient use.
Lesson: A pass with a dummy plug only proves part of the chain — always finish with the real load connected, and when a fault shows up only under load, suspect aged, cracked solder joints first.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.