Siemens Magnetom C!: empty raw data disks on the imager
What happened
Raw data was missing on the imager of an MRI scanner: the raw data disks were empty and the data had to be recreated manually before scanning could resume.
What was checked
- Tried a service batch script for recreating raw data; per the case, its name and location vary with the software version, and it did not solve the problem on its own.
- Deleted the raw data on the reconstructor and rebooted it to force a rebuild; this also did not restore a working state.
- Checked the disk space configuration: the case reported four disks, while the documentation seemed ambiguous about the layout, and another similar unit was reported with three disks (one for the OS, two in RAID).
Cause
No definite root cause was established in the case: the engineers found the imager's raw data area empty, and the available on-line rebuild attempts did not bring the storage back to a consistent state.
What was done
The engineers restored the imager host from an older saved image of a known working system (reported in the case as about two years old); after the restore the raw data area was recreated and the scanner returned to operation.
Safety
The main magnet of an MRI is always energised, so during any disk or image-restore work keep ferromagnetic tools and spare drives away from the bore, and before return to clinical use complete the manufacturer's electrical safety checks.
Lesson: Keep a current, verified backup image of the host system after every software change, because scripted rebuild attempts may fail and the image is the fastest way back to a working scanner.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.