Siemens · Magnetom C!

Siemens Magnetom C!: empty raw data disks on the imager

What happened

Raw data was missing on the imager of an MRI scanner: the raw data disks were empty and the data had to be recreated manually before scanning could resume.

What was checked

Cause

No definite root cause was established in the case: the engineers found the imager's raw data area empty, and the available on-line rebuild attempts did not bring the storage back to a consistent state.

What was done

The engineers restored the imager host from an older saved image of a known working system (reported in the case as about two years old); after the restore the raw data area was recreated and the scanner returned to operation.

Safety

The main magnet of an MRI is always energised, so during any disk or image-restore work keep ferromagnetic tools and spare drives away from the bore, and before return to clinical use complete the manufacturer's electrical safety checks.

Lesson: Keep a current, verified backup image of the host system after every software change, because scripted rebuild attempts may fail and the image is the fastest way back to a working scanner.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.