Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Siemens Magnetom Concerto 0.2T
Siemens · Magnetom Concerto 0.2T

Siemens Magnetom Concerto 0.2T: typical weak points during servicing

What happened

Not a specific fault: an engineer taking over this low-field open MRI asked what usually fails on this model and what to focus on at service.

What was checked

Cause

No single cause in this case: the listed items are known age-related weak spots on this model — fouling of the water loop, drift of power-supply voltages, worn connector contacts and wear of the table brake components.

The cause is probable and was not fully confirmed.

What was done

The engineers shared a preventive checklist rather than a repair: keep the Lambda supply voltages within nominal, clean the water strainer on schedule, watch the refrigerant circuit and pump performance, inspect and maintain the coil preamplifier plugs (in this case, with one plug faulty, operation on the remaining good plug was the workaround reported), and replace worn table parts — brake cables, encoder belt and Z-brake rubbers — before they fail. The benefit of these measures is probable, as this is preventive advice from one case, not a confirmed repair.

Safety

The magnetic field is always present and the cabinets store energy: keep ferromagnetic tools and parts under control in the scan room, and before any work inside the ECA cabinet isolate, lock out and tag the power and verify discharge of the supplies.

Lesson: When you take over a scanner model that is new to you, build the preventive maintenance plan from known weak points — cooling hygiene, power-supply voltages, connectors and table mechanics — instead of waiting for failures.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes