Siemens Magnetom Essenza: intermittent coil-related system errors

What happened

The scanner periodically stopped with coil-related errors and asked for an engineer to be called.

What was checked

Cause

In this case the engineers found poor contact in the coil connectors — on the coils themselves and in the table — together with a damaged head coil connector.

What was done

Contacts on the coils and in the table were cleaned with a suitable contact cleaner, and the damaged head coil connector was repaired. The team also stopped repeated plugging and unplugging: the head coil stays connected. In this case the errors stopped after this work.

Safety

Work near the bore under MR zone discipline — no ferromagnetic tools or loose items — keep cleaning fluid off open contacts and electronics, and confirm normal system behavior before the scanner returns to patient use.

Lesson: Intermittent coil errors on an MRI should be fought at the connectors first — worn contacts and needless reconnection cycles explain more faults than electronics do.