Siemens Magnetom Essenza: intermittent coil-related system errors
What happened
The scanner periodically stopped with coil-related errors and asked for an engineer to be called.
What was checked
- Inspect the coil connectors and the mating connectors in the patient table for wear, bent contacts and play.
- Clean the contact surfaces on the coils and in the table connectors.
- Boot the system without the local coils to separate a head coil fault from local coil faults.
Cause
In this case the engineers found poor contact in the coil connectors — on the coils themselves and in the table — together with a damaged head coil connector.
What was done
Contacts on the coils and in the table were cleaned with a suitable contact cleaner, and the damaged head coil connector was repaired. The team also stopped repeated plugging and unplugging: the head coil stays connected. In this case the errors stopped after this work.
Safety
Work near the bore under MR zone discipline — no ferromagnetic tools or loose items — keep cleaning fluid off open contacts and electronics, and confirm normal system behavior before the scanner returns to patient use.
Lesson: Intermittent coil errors on an MRI should be fought at the connectors first — worn contacts and needless reconnection cycles explain more faults than electronics do.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.