Siemens · Magnetom Essenza

Siemens Magnetom Essenza: patient table will not drive, false 'patient moved' message

What happened

After the patient was fed into the bore, the table refused to move under program control and reported that the patient had moved. Restarting the table did not help.

What was checked

Cause

In this case the engineers found two drive boards failing in sequence: first the horizontal movement board, and then, after it was replaced, the vertical movement board. The 'patient moved' message was a false flag from the bad boards, not real patient motion.

What was done

The horizontal drive board was replaced; when the vertical drive then failed, that board was replaced too. After both board swaps the table moved normally under program control, and the case was closed as solved.

Safety

Before opening the drive units, isolate the table supply and secure the table against rolling — with the brake released it can move freely — and carry out the electrical safety test before returning the table to patient use.

Lesson: A motion warning that blames the patient can originate in the drive electronics itself, so test the drive boards one by one before chasing motors, belts and mechanics.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.