Siemens · Magnetom Multiva 1.5T

Siemens Magnetom Multiva 1.5T: patient table creeps out and will not release

What happened

The patient table crept out slowly and would not release; the fault was intermittent and sometimes settled temporarily while the control joystick was being operated.

What was checked

Cause

The engineers found a faulty optocoupler inside the control joystick: its drifting output was misread by the table control as movement and release commands, which matches the creeping table and the floating error pattern reported in the case.

What was done

The faulty optocoupler in the joystick was replaced. After the replacement the table moved and released normally and the fault did not return — the case was reported as solved.

Safety

Before any work on the table or joystick, take the patient off the table and secure it against unexpected movement, use only non-magnetic tools inside the magnet room, and function-test the stop and release controls before returning the table to use.

Lesson: When an intermittent table fault reacts to handling the controller, check the controller's own electronics before dismantling the actuator.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.