Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Siemens Magnetom Prisma Fit
Siemens · Magnetom Prisma Fit

Siemens Magnetom Prisma Fit: doubtful helium level reading after sensor replacement

What happened

After the helium level sensor (MSUP) was replaced with a unit taken from another scanner of a related model, the display showed about 974 liters (45%) as reported in the case, while the vessel capacity on the magnet nameplate was 2160 liters. The engineers suspected the reading was incorrect.

What was checked

Cause

In this case there was no fault: the sensor calibration refers to the working fill volume, reported as about 1200 liters of liquid helium, while the 2160 liters on the nameplate is the total vessel capacity. The displayed 974 liters therefore corresponds to roughly 70% full.

What was done

No repair was needed. The reading was accepted as correct once the level was interpreted against the actual working fill volume rather than the total vessel capacity, and the engineers continued tracking the level against this reference.

Safety

Liquid helium is an extreme cryogen that can cause severe cold burns and displace oxygen in enclosed spaces, so follow the site cryogen procedures around the magnet and remember that the magnetic field is always present.

Lesson: A helium level reading can look wrong simply because the nameplate capacity and the calibrated fill volume differ, so compare like with like before suspecting the sensor.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes