Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Siemens Magnetom (мануал MPS 3600, упоминается Symphony)
Siemens · Magnetom (мануал MPS 3600, упоминается Symphony)

Siemens Magnetom: switch heater energised only at start and end of ramp-up

What happened

Engineers asked why, during ramp-up (Run to Field) on a Magnetom superconducting magnet, the switch heater (SH) is energised only at the very start of the current ramp and again near the end at stabilisation, instead of staying on through the whole ramp. The discussion also covered two quenches of a Symphony magnet during ramp-up driven by a third-party power supply.

What was checked

Cause

The engineers found that during the ramp the coil's inductive voltage appears across the switch, so a tiny current through the high-resistance switch — microamps, as discussed in the case — keeps it warm on its own. Once the magnet reaches its working persistent current, this voltage and the switch current drop to zero, and the switch then needs external heating. That is why SH is switched on at the start and again near the end of the ramp, not throughout it.

What was done

The correct sequence follows the manufacturer's MPS 3600 documentation — a procedure from the manufacturer's service documentation, done by authorised personnel: SH energised at the start of the ramp, de-energised once the circuit current is established, re-energised shortly before the nominal current, and de-energised again after the overshoot and stabilisation (in the case's example, off by about 10 A and back on roughly 10 A below nominal). In Run to Zero, SH is switched on again once the magnet voltage falls below about one volt, as reported in the case.

Safety

A superconducting magnet stores enormous magnetic energy, and a mistake during ramping can trigger a quench with rapid cryogen release — such work must be left to authorised personnel with the quench protection verified and the magnet voltage monitored continuously.

Lesson: Understand the physics behind each step of a magnet procedure before changing it — a heater that switches at seemingly odd moments is often doing exactly what it should, and improvising during ramp-up has already cost magnets.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes