Siemens Magnetom Symphony: scanner fails to start in the morning, no gradients
What happened
As reported in the case, every morning start failed: the BTB breaker in the gradient cabinet tripped, there were no gradients, and the RCA1 error came up. After several reboots — with the gradient circuit responding to tapping — the system then ran normally for the rest of the day.
What was checked
- Review the MPCU logs to see which gradient-cabinet module reports the fault first at power-up; in this case the logs pointed to the MSUP.
- Check the history of the modules in the cabinet: a shelf known to have been moved from another system is a prime suspect for intermittent, hard-to-repeat faults.
Cause
The engineers found that the MSUP module was faulty; moreover, the whole MSUP shelf was not original to this scanner — it had been installed from another system, which explains the intermittent nature of the fault.
What was done
The MSUP shelf was replaced as a whole unit. After that the morning starts returned to normal and the error did not recur; the outcome is reported as solved.
Safety
Gradient amplifier cabinets contain capacitors that store lethal energy after power-off: never open or tap on the cabinet while it is energised — wait out the manufacturer's discharge interval, verify absence of voltage, and leave internal work to authorised personnel.
Lesson: An intermittent fault that clears after reboots or tapping must be traced in the logs to a specific module, and modules transplanted from another system are a classic hidden cause.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.