Siemens Magnetom Symphony: scanner fails to start in the morning, no gradients

What happened

As reported in the case, every morning start failed: the BTB breaker in the gradient cabinet tripped, there were no gradients, and the RCA1 error came up. After several reboots — with the gradient circuit responding to tapping — the system then ran normally for the rest of the day.

What was checked

Cause

The engineers found that the MSUP module was faulty; moreover, the whole MSUP shelf was not original to this scanner — it had been installed from another system, which explains the intermittent nature of the fault.

What was done

The MSUP shelf was replaced as a whole unit. After that the morning starts returned to normal and the error did not recur; the outcome is reported as solved.

Safety

Gradient amplifier cabinets contain capacitors that store lethal energy after power-off: never open or tap on the cabinet while it is energised — wait out the manufacturer's discharge interval, verify absence of voltage, and leave internal work to authorised personnel.

Lesson: An intermittent fault that clears after reboots or tapping must be traced in the logs to a specific module, and modules transplanted from another system are a classic hidden cause.