Siemens Optima: control pedal failure

What happened

The control pedal of the angiography system stopped working. A pedal from an INNOVA unit was tried as a substitute but did not fit the Optima, as reported in the case.

What was checked

Cause

Broken or damaged wires inside the pedal, hidden under the potting compound.

What was done

The engineer removed the potting compound, cut out the damaged section of the cable — about 10 cm, as reported in the case — and re-soldered the wires. The pedal worked again and the system was back in service. Note from the case: a pedal from an INNOVA unit is not a direct replacement for the Optima.

Safety

After opening the pedal and re-soldering, run the electrical safety test (ground continuity and leakage) and verify full pedal function before returning the system to clinical use.

Lesson: A pedal from a related model is not a drop-in substitute, so verify compatibility first, and remember that a simple wiring fault can hide under the potting compound.

Covered in the course: Angiography and C-arm Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.