Siemens Optima: control pedal failure
What happened
The control pedal of the angiography system stopped working. A pedal from an INNOVA unit was tried as a substitute but did not fit the Optima, as reported in the case.
What was checked
- Tried a pedal from an INNOVA unit on the Optima — not a direct fit (as reported in the case).
- Opened the pedal housing and removed the potting compound; found a simple PCB underneath.
- Inspected the internal wiring and suspected an open circuit or a short.
Cause
Broken or damaged wires inside the pedal, hidden under the potting compound.
What was done
The engineer removed the potting compound, cut out the damaged section of the cable — about 10 cm, as reported in the case — and re-soldered the wires. The pedal worked again and the system was back in service. Note from the case: a pedal from an INNOVA unit is not a direct replacement for the Optima.
Safety
After opening the pedal and re-soldering, run the electrical safety test (ground continuity and leakage) and verify full pedal function before returning the system to clinical use.
Lesson: A pedal from a related model is not a drop-in substitute, so verify compatibility first, and remember that a simple wiring fault can hide under the potting compound.
Covered in the course: Angiography and C-arm Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.