Siemens RapidPoint 500: cartridge failures in a row, pO2 stops measuring

What happened

Cartridges failed one after another: after roughly three weeks onboard, pO2 readings stopped and the analyzer refused to run measurements (as reported in the case).

What was checked

Cause

In this case, operators were not seating the cartridge fully home, and the analyzer-side contact group bent; every cartridge inserted afterwards was damaged on contact, so three failed in a row (as reported). Separately, the pO2 channel degraded after about three weeks of onboard storage, matching the case report.

What was done

The installation practice was corrected: the cartridge must be inserted and pressed fully home until it locks, and the contact area was inspected before new cartridges were installed. After that no further cartridge failures were reported; the site later chose to move this workload to a different blood gas platform.

Safety

Power the analyzer down and disconnect it from mains before inspecting the cartridge bay or contact group, and treat all cartridges and spilled sample fluid as potentially infectious — wear gloves and disinfect the work area.

Lesson: When consumables fail one after another, check the mating contacts and how staff seat the cartridge before ordering more replacements.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.