Siemens S2000: fans cycle on and off at power-up

What happened

The AC OK LED on the power supply stayed lit. When mains was applied, the fans started and stopped in a repeating cycle. When the small 2-pin connector feeding the VI unit was unplugged, the cycling stopped and the behaviour became stable.

What was checked

Cause

The engineers found that the cycling disappeared only when the VI load was disconnected, which pointed to a power supply unable to hold its outputs stable under load. The supply was judged faulty; this conclusion was probable and was not confirmed by a completed repair in the case.

The cause is probable and was not fully confirmed.

What was done

Diagnose and repair or replace the power supply, following the manufacturer's service documentation. No repair was completed in the case, so the conclusion remains probable.

Safety

A switched-mode power supply keeps a dangerous charge on its capacitors even after the plug is out: isolate from mains, let the capacitors discharge, verify the absence of voltage before working inside, and run the electrical safety test before the scanner returns to use.

Lesson: When a fault appears or disappears with one unit unplugged, use that load step by step to localise the problem, but prove the fault at the supply itself before condemning it.