Siemens Somatom Confidence: images not auto-transferring to archiving uploader

What happened

A Somatom Confidence CT had no automatic route to the uploader computer that archives the studies: after each scan the images stayed on the scanner. The task was to make the CT push every study to the uploader automatically.

What was checked

Cause

No hardware or software fault was found. The destination had simply never been configured: on this scanner, automatic image routing is defined in the network configuration as a host and a destination node, and the node must then be assigned to each scan protocol that should send automatically — a protocol without the assigned node keeps its images local.

What was done

The change was made as a configuration procedure from the manufacturer's service documentation, done by authorised personnel: the uploader was registered as a host in the scanner's service-level network configuration, a destination network node was created for it, and that node was assigned to every scan protocol. As reported in the case, after this the CT sent each study to the uploader automatically and the issue was closed.

Safety

Because the change touches scan protocols, run a test acquisition and confirm that no imaging or dose-related parameters were altered before the scanner is returned to clinical use.

Lesson: Automatic image routing is a per-protocol setting: a new archiving destination only works once its host is registered on the scanner and the destination node is attached to every protocol that should send.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.