Siemens Somatom Definition: PACS send delivers only one image per study
What happened
Reported in the case: sending studies to PACS was unreliable — some studies went out with only a single image instead of the full DICOM series.
What was checked
- Rule out the link first: confirm the send configuration and destination are unchanged and other studies still reach PACS — this separates a network problem from a local one.
- Inspect the local DICOM database for the affected studies; look for corrupt or incomplete series entries where the send stops after the first image.
- Do a test send of a small known-good series before and after any intervention, so you can prove what actually fixed it.
Cause
No root cause was documented, but the engineers found the fault disappeared only after a full database rebuild, so they concluded the local DICOM databases had become corrupt or inconsistent — single images slipped through while full series failed.
What was done
Before deleting anything, the engineers exported the needed series as ISO images through the file section into the service extraction folder. Then they performed a full wipe of the local databases — a procedure from the manufacturer's service documentation, done by authorised personnel. After the rebuild, full series went to PACS normally and the fault was solved.
Safety
A database wipe is destructive: before you start, make sure every needed study is exported and verifiably readable, and confirm with a test send that data reaches PACS before the scanner returns to clinical use.
Lesson: Always extract the affected studies to storage you control before any destructive database recovery — a verified backup turns a risky wipe into a routine repair.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.