Siemens Somatom go.Up: no automatic sending to PACS after interface setup

What happened

After the interface for a new PACS server was configured, the scanner did not send studies automatically, although the send settings window looked correct.

What was checked

Cause

Automatic forwarding depends on per-protocol send rules in the protocol editor, not only on the destination entered in the interface settings; since the new PACS had not been added to those rules, no study was matched for an automatic send.

What was done

The engineers added the new PACS server to the send rules in the protocol editor and confirmed on a study that automatic forwarding to it now worked; the issue was resolved.

Safety

The work is settings-only, with no radiation or electrical exposure, but make the change between studies, never during an active scan, and verify with a test send before returning the scanner to clinical use so no patient images silently fail to reach the archive.

Lesson: A new PACS destination must be set not only in the interface settings but also in the send rules inside the protocol editor — otherwise no study matches an automatic send.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.