Siemens SOMATOM Sensation 64: images not sent to PACS after workstation backup restore
What happened
After a backup from the previous workstation was restored onto the replacement station, the CT acquired and reconstructed normally, but images stopped reaching PACS even though the PACS host answered ping.
What was checked
- Separated connectivity from DICOM: the PACS host answered ping, so the network was up and the fault had to be one layer above it.
- Recreated the DICOM send rule and ran verification from the scanner side.
- Checked verification from both directions, including what the PACS itself reported, to rule out a mismatch at the archive.
Cause
Restoring a backup made for the older workstation onto different hardware had knocked out the PACS adapter settings. Acquisition and reconstruction were untouched, but the adapter had lost the configuration it needs to push studies out, so sends failed while ping kept working.
What was done
The engineers rebuilt the adapter configuration, after which image transfer to PACS worked again. Recovering a workstation from a backup onto other hardware is a procedure from the manufacturer's service documentation, done by authorised personnel — not a straight file copy.
Safety
De-energise the workstation at the mains before touching its rear connections, and before returning it to clinical use confirm that studies reach the correct destinations and that no images were lost or misrouted while the fault was active.
Lesson: A working ping proves nothing above the network layer — after any workstation recovery or backup restore, verify the whole DICOM send path end-to-end, in both directions, before releasing the system.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.