Siemens · Symphony 1.5T

Siemens Symphony 1.5T: adjust fails while all units show ready

What happened

System adjust did not complete in either automatic or manual mode, yet every unit reported ready with green indicators; no error codes were reported in the case.

What was checked

Cause

The engineers found faulty optics in the fibre-optic communication channel on the reconstructor side; the units kept reporting ready because their status signals did not reflect the failed link.

What was done

The engineers replaced the optical module in the fibre line on the reconstructor; after the replacement the adjust passed and the system returned to normal operation.

Safety

Before opening the reconstruction cabinet, isolate and lock off the AC supply and allow internal circuits to discharge, and confirm the system passes the manufacturer's safety checks before return to clinical use.

Lesson: Green ready indications prove only that units power up and self-report, not that the data path between them is healthy, so when adjust fails without error codes, test the communication links such as fibre optics before suspecting the coil or the adjust routine itself.