Siemens Symphony / Avanto: chiller refrigerant top-up with fouled condensers
What happened
The engineers found the compressor-condenser unit of the MRI cooling system low on refrigerant and needing a top-up; the condenser coils were heavily fouled after years without cleaning, as described in the case.
What was checked
- Inspect the condenser coils first — in this case years of dust had built up and restricted airflow through them.
- Record the refrigerant circuit pressures and assess the charge before adding anything.
- If low charge recurs, search for a leak (oil traces at joints and fittings) instead of simply refilling again.
Cause
Reported in the case: gradual loss of refrigerant charge over time, combined with condenser coils that had not been cleaned for years, which reduced heat rejection and added load to the compressor.
What was done
The engineers topped up the refrigerant through the charging port using a refill device — reportedly about five minutes of work — and washed the condenser coils with a pressure washer; even after years without cleaning the coils came out clean. Refrigerant work was done by trained personnel following the manufacturer's service documentation. The case outcome was reported as solved.
Safety
De-energise and lock off the compressor-condenser unit before washing the coils, keep water clear of electrical enclosures, fan motors and wiring, and let everything dry and pass an electrical safety check before the unit is returned to use.
Lesson: Recurring loss of refrigerant charge means there is a leak to find, and regular condenser cleaning is the cheapest protection the compressor gets.