Siemens · Symphony / Avanto

Siemens Symphony / Avanto: chiller refrigerant top-up with fouled condensers

What happened

The engineers found the compressor-condenser unit of the MRI cooling system low on refrigerant and needing a top-up; the condenser coils were heavily fouled after years without cleaning, as described in the case.

What was checked

Cause

Reported in the case: gradual loss of refrigerant charge over time, combined with condenser coils that had not been cleaned for years, which reduced heat rejection and added load to the compressor.

What was done

The engineers topped up the refrigerant through the charging port using a refill device — reportedly about five minutes of work — and washed the condenser coils with a pressure washer; even after years without cleaning the coils came out clean. Refrigerant work was done by trained personnel following the manufacturer's service documentation. The case outcome was reported as solved.

Safety

De-energise and lock off the compressor-condenser unit before washing the coils, keep water clear of electrical enclosures, fan motors and wiring, and let everything dry and pass an electrical safety check before the unit is returned to use.

Lesson: Recurring loss of refrigerant charge means there is a leak to find, and regular condenser cleaning is the cheapest protection the compressor gets.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.