Siemens Symphony: body coil overheat errors with normal cooling flow
What happened
The system repeatedly flagged the body coil as overheating, driven by a 'no cooling flow' signal, although the actual cooling flow was normal.
What was checked
- Reported in the case: the engineers verified the real cooling flow and found it normal, so the alarm pointed to the sensing side rather than the cooling circuit.
- Reported in the case: diagnostics of the flow sensor raised the suspicion that the sensor (impeller) had failed while the flow was still present.
Cause
In this one case the engineers found the most likely cause to be a failed flow sensor (impeller) in the body coil cooling circuit, producing a false 'no flow' signal while the real flow was normal.
The cause is probable and was not fully confirmed.
What was done
The workaround ideas recorded in the case emulate the flow signal and therefore disable the coil overheat protection — they are not a repair and must not be used. The correct fix here is a procedure from the manufacturer's service documentation, done by authorised personnel, replacing the faulty sensor so the alarm again reflects the real flow.
Safety
The body coil overheat protection exists to prevent patient burns — never return the system to scanning with the flow signal emulated or the alarm bypassed; confirm the sensor and the whole protection chain work before the next patient.
Lesson: A 'no flow' alarm with normal actual flow means the sensor has failed safe — replace the sensor to restore honest monitoring instead of faking the signal.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.