Siemens Symphony: cooling fan failure over the active shim
What happened
A cooling fan above the active shim had failed and stopped running.
What was checked
- Worked out the fan parameters needed for a replacement — supply voltage and frame size; reported in the case as 220 V AC and 119x119x38 mm.
- Checked the old fan electrically — the engineers found an open circuit in its motor winding.
Cause
An open circuit in the motor winding of the old fan that had reached the end of its service life.
What was done
The engineers fitted a suitable 220 V AC axial fan of matching frame size, sourced from an electronics parts shop; per the case it ran for a couple of years afterwards. The model named in the case is 4715FS-23T-B50, 119x119x38 mm, 220 V AC.
Safety
Isolate the fan's mains supply and confirm zero voltage before touching the wiring, and follow magnet-room discipline: only non-magnetic tools and parts may be taken into the strong static magnetic field.
Lesson: When cooling in the shim area is in question, test the fan itself first — a simple winding check often confirms the fault, and supply voltage plus frame size are the key figures when sourcing a replacement.