Academy / Field fault notes / Siemens Symphony
Siemens · Symphony

Siemens Symphony: cooling fan failure over the active shim

What happened

A cooling fan above the active shim had failed and stopped running.

What was checked

Cause

An open circuit in the motor winding of the old fan that had reached the end of its service life.

What was done

The engineers fitted a suitable 220 V AC axial fan of matching frame size, sourced from an electronics parts shop; per the case it ran for a couple of years afterwards. The model named in the case is 4715FS-23T-B50, 119x119x38 mm, 220 V AC.

Safety

Isolate the fan's mains supply and confirm zero voltage before touching the wiring, and follow magnet-room discipline: only non-magnetic tools and parts may be taken into the strong static magnetic field.

Lesson: When cooling in the shim area is in question, test the fan itself first — a simple winding check often confirms the fault, and supply voltage plus frame size are the key figures when sourcing a replacement.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.