Siemens Symphony/Harmony: cryoshield temperature measured only with an external meter
What happened
The engineers needed to read the temperature of the cryogenic shields (the barrel) with an external instrument and were looking for a proven schematic of a measuring adapter, because the handy multimeter methods did not agree with the host readings.
What was checked
- As reported in the case, the engineers fed a 100 µA current through the pn junction of the transistor sensor inside the shield barrel and measured the forward voltage drop with a multimeter.
- They also tried a Fluke meter in diode-test mode; its readings came out roughly 10 K more optimistic than the host values, as reported in the case.
Cause
No hardware fault was identified in the case. The shield sensors are pn-junction transistors that need a defined, stable test current; a handheld meter's diode-test mode drives a different current, which plausibly explains the roughly 10 K discrepancy noted in the case.
What was done
The engineers built a small external adapter: a 100 µA current source on a single IC using a Howland topology, similar to a circuit they found on another manufacturer's MRI board. With the adapter any multimeter could read the junction sensor, and the case was closed as solved. The manufacturer's service documentation also describes the diode-test-mode measurement as an alternative method.
Safety
Treat the magnet room as a high-field area — use only non-magnetic, MR-approved tools and instruments, and never probe the cryostat's cold penetrations without the manufacturer's procedure, since the shields sit at cryogenic temperatures and mishandling can lead to a quench.
Lesson: A junction temperature sensor is only as accurate as the test current you drive it with, so build or use a defined, repeatable current source and verify it against the host readings before trusting any external measurement.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.