Siemens Symphony: host and reconstruction software recovery
What happened
A question raised in one case: how to rebuild the host and reconstruction station software, with a concern that disks behind the RAID controller would not be visible to third-party disk-imaging software.
What was checked
- Recovery of the host was attempted from a previously stored full-disk image made with third-party imaging software.
- The reconstruction station was installed from scratch using the original base-system installation media supplied by the manufacturer, which partitions the disks itself.
Cause
No hardware failure was reported in the case; it was a software rebuild after loss of the installed system, so no specific fault cause was identified.
What was done
As reported in the case: the host was restored from the stored full-disk image, and the imaging tool saw the RAID-attached disks without problems. The reconstruction station was rebuilt from scratch from the original installation media, which partitions and prepares the disks itself. The engineers noted that without a stored image the reconstruction software will not simply transfer to a new disk set — the original installation media is required.
Safety
After any software rebuild, keep the scanner out of clinical use until the manufacturer's verification procedures and the electrical safety test are passed — and remember the magnet field is always on, so ferromagnetic-safety rules apply even for work in the computer room.
Lesson: Keep a verified full-disk image of the host and the original installation media for the reconstruction station ready before a software recovery is ever needed.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.