Siemens Symphony: hunting an electromagnetic interference source
What happened
The engineers were chasing an electromagnetic interference source and wanted a way to software-exclude the patient table unit (PTAB) to see whether the noise disappeared.
What was checked
- Tried switching the PTAB off via its CAN link (the optical/laser line), but as reported in the case this only takes effect together with the GPA.
- Physically disconnected the PMU as the next suspect and watched whether the interference changed.
Cause
The engineers found that the PMU itself was emitting the interference; the PTAB turned out not to be the source in this case.
What was done
Disconnecting the PMU removed the interference, which confirmed it as the source. The idea of software-emulating or excluding the PTAB for such tests was left unresolved in this case.
Safety
Any disconnection or reconnection of MRI subsystem electronics must follow the manufacturer's power-down and isolation procedure, and the required electrical safety tests from the service documentation must be passed before the system returns to clinical use.
Lesson: When a software exclusion for a suspected interference source is not available, a controlled physical disconnection of the suspect unit, done per the documentation, will still identify the culprit.