Academy / Field fault notes / Siemens Symphony
Siemens · Symphony

Siemens Symphony: hunting an electromagnetic interference source

What happened

The engineers were chasing an electromagnetic interference source and wanted a way to software-exclude the patient table unit (PTAB) to see whether the noise disappeared.

What was checked

Cause

The engineers found that the PMU itself was emitting the interference; the PTAB turned out not to be the source in this case.

What was done

Disconnecting the PMU removed the interference, which confirmed it as the source. The idea of software-emulating or excluding the PTAB for such tests was left unresolved in this case.

Safety

Any disconnection or reconnection of MRI subsystem electronics must follow the manufacturer's power-down and isolation procedure, and the required electrical safety tests from the service documentation must be passed before the system returns to clinical use.

Lesson: When a software exclusion for a suspected interference source is not available, a controlled physical disconnection of the suspect unit, done per the documentation, will still identify the culprit.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.