Siemens Symphony: image artifacts with normal QA

What happened

Artifacts on images on a Siemens Symphony while QA and the standard test tools report normal results. The engineers suspected the gradient system or the proton frequency.

What was checked

Cause

In this case the cause is reported as probable: loosened fastening of the gradient cables at the cable-to-busbar transition on the gradient cylinder. Vibration and thermal cycling can work such hardware loose over time, and the resulting micro-movement of the cables can show up as image artifacts even when QA and test tools stay within limits — a similar case was resolved exactly this way.

The cause is probable and was not fully confirmed.

What was done

The fix in the similar case was to re-tighten and restore the fastening of the gradient cables at the cable-to-busbar transition, working to the torques and layout given in the manufacturer's service documentation. Since the outcome here is probable rather than confirmed, verify the result with a full quality check before returning the system to clinical use.

Safety

De-energise and lock out the gradient chain before touching any cable run — it carries high peak currents and stored energy — and keep magnet-room discipline throughout: no ferromagnetic tools or loose hardware anywhere near the bore.

Lesson: When images degrade but QA stays clean, look at mechanical integrity first — loose gradient cable hardware is a classic hidden source of artifacts.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.