Siemens · Symphony

Siemens Symphony: Patient table dead with PTAB communication fault

What happened

The patient table would not power up, and the system pointed to a PTAB communication fault on the table.

What was checked

Cause

The engineers found poor contact on the pins of the joystick connector on the PTAB backplane: the solder joint no longer held reliably and broke the communication path.

What was done

The joystick connector on the PTAB backplane was re-soldered, restoring solid contact; after that the table powered up normally and, as reported in the case, the fault did not return.

Safety

De-energise the table and verify absence of voltage before soldering on the backplane, and keep every tool accounted for — the MRI magnet field is always on, so a loose ferrous item near the bore is a serious hazard.

Lesson: A communication fault that makes an assembly look dead can come down to a single connector or solder joint, so check contacts and interconnects before ordering boards.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.