Siemens Symphony: patient table motion faults stopping scans

What happened

Scan aborted mid-run on FLAIR with a patient-table error; on diffusion sequences the table started moving on its own.

What was checked

Cause

The engineers found two contributors in this case: chatter of worn microswitches on the table's pinch-protection plate, showing up only on high-gradient sequences because of vibration, and a cold solder joint in the joystick contacts sending spurious motion commands. A tired joystick centering spring was also reported in the case as another possible source of the same symptoms.

What was done

Restored the solder joints on the joystick contacts and eliminated the microswitch chatter on the pinch-protection plate. The table motion faults did not return; the problem was solved in this case.

Safety

Before returning the table to use, test the pinch-protection plate, the emergency stop and the table motion controls under real load, since self-driven table motion with a patient aboard is a direct crush hazard.

Lesson: Table motion errors that appear only on high-gradient sequences are usually simple contact problems — switch chatter or bad solder joints — so check the switches and joystick wiring before suspecting the drive electronics.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.