Siemens Symphony: patient table moves in jerks during vertical travel

What happened

When moving the table up or down, it travels a short distance and stops even with the joystick held; pressing the joystick again repeats the same cycle. No error codes were reported in the case.

What was checked

Cause

In this case the engineers considered an exhausted grease cartridge the likely cause: the vertical-drive screw was running nearly dry, so the drive stalled intermittently under load. This was a probable diagnosis in one case, not a confirmed rule.

The cause is probable and was not fully confirmed.

What was done

Replace the table grease cartridge and, if needed, apply a moderate amount of grease to the screw manually, following the manufacturer's service manual. Also inspect the brake at the bottom of the column: the engineers warned that with a broken brake the table can drift down under a heavy patient even after the cartridge is replaced.

Safety

Before returning the table to patient use, verify that the vertical-drive brake holds the load at any height with the drive de-energised — a table that can sink under a patient is a direct crush hazard.

Lesson: Jerky, stop-and-go motion on a motorised patient table usually points to lubrication starvation first, so check the grease supply and the load-holding brake before digging into the drive electronics.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.