Siemens Symphony: patient table moves in jerks during vertical travel
What happened
When moving the table up or down, it travels a short distance and stops even with the joystick held; pressing the joystick again repeats the same cycle. No error codes were reported in the case.
What was checked
- Established how long it had been since the last maintenance and greasing of the patient table.
- Inspected the grease cartridge and the condition of the lubrication supply to the vertical-drive screw.
Cause
In this case the engineers considered an exhausted grease cartridge the likely cause: the vertical-drive screw was running nearly dry, so the drive stalled intermittently under load. This was a probable diagnosis in one case, not a confirmed rule.
The cause is probable and was not fully confirmed.
What was done
Replace the table grease cartridge and, if needed, apply a moderate amount of grease to the screw manually, following the manufacturer's service manual. Also inspect the brake at the bottom of the column: the engineers warned that with a broken brake the table can drift down under a heavy patient even after the cartridge is replaced.
Safety
Before returning the table to patient use, verify that the vertical-drive brake holds the load at any height with the drive de-energised — a table that can sink under a patient is a direct crush hazard.
Lesson: Jerky, stop-and-go motion on a motorised patient table usually points to lubrication starvation first, so check the grease supply and the load-holding brake before digging into the drive electronics.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.