Siemens Symphony: poor image quality on foot scans

What happened

Radiologists were not satisfied with the quality of foot studies, while knee studies performed on the same knee coil drew no complaints. No error codes were reported.

What was checked

Cause

The engineers found that the poor image quality was aliasing: with the knee coil and its small field of view, tissue outside the scanned volume wrapped onto the image of the foot. Coil selection and positioning did not match the anatomy being examined.

What was done

The recommended fix was to use a Flex coil or the head coil for the foot and suppress any remaining wrap with phase oversampling or saturation bands. If the knee coil had to be kept, a prone position with the patient clearly told which part is being imaged, together with oversampling/saturation, gave acceptable quality. The case was reported as solved.

Safety

Before scanning, check that coil cables run without tight loops and are padded away from the patient's skin, since RF heating at cable contact points can cause burns.

Lesson: Match the coil, field of view and positioning to the anatomy actually being imaged — a quality complaint may be aliasing, not a failing coil.

Covered in the course: MRI Service — Level 2: Fault-Finding →