Siemens Symphony: poor image quality on foot scans
What happened
Radiologists were not satisfied with the quality of foot studies, while knee studies performed on the same knee coil drew no complaints. No error codes were reported.
What was checked
- Review the foot images for aliasing (wrap-around) over the region of interest.
- Compare the selected coil and scan field of view with the anatomy actually being imaged.
- Check whether phase oversampling or a saturation band removes the wrap.
- Verify the patient positioning that was used with the knee coil.
Cause
The engineers found that the poor image quality was aliasing: with the knee coil and its small field of view, tissue outside the scanned volume wrapped onto the image of the foot. Coil selection and positioning did not match the anatomy being examined.
What was done
The recommended fix was to use a Flex coil or the head coil for the foot and suppress any remaining wrap with phase oversampling or saturation bands. If the knee coil had to be kept, a prone position with the patient clearly told which part is being imaged, together with oversampling/saturation, gave acceptable quality. The case was reported as solved.
Safety
Before scanning, check that coil cables run without tight loops and are padded away from the patient's skin, since RF heating at cable contact points can cause burns.
Lesson: Match the coil, field of view and positioning to the anatomy actually being imaged — a quality complaint may be aliasing, not a failing coil.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.