Siemens Symphony (RCA) MRI: pump circuit current limiter replacement
What happened
A current-limiting device in the pump power circuit of a Siemens Symphony (RCA) MRI needed replacement, but the component first had to be identified before a substitute could be found.
What was checked
- The engineers identified the device in the pump starter circuit: the 'current limiter' turned out to be a standard thermal (overload) relay.
- The replacement was matched to the existing contactor: same form factor and current setting, plus a separate coil (reported in the case as 24 V).
- A three-pole circuit breaker after the contactor was considered as an alternative, accepting the loss of auxiliary contacts and the adjustable trip setting.
Cause
The case did not report why the relay had to be replaced; the practical difficulty was identification — a part labelled on the pump circuit as a 'current limiter' is, in fact, a common thermal overload relay.
What was done
Replace the relay with a thermal overload unit of the same form factor and current setting — the case mentions a standard IEK relay with a separate contactor coil (reported as 24 V) or any suitable analogue. A three-pole circuit breaker after the contactor was noted as a workable alternative where auxiliary contacts and an adjustable setting are not needed. Set the overload to the pump motor's rated current; final part selection per the manufacturer's service documentation. With the component identified, the sourcing question in this case was resolved.
Safety
Isolate and lock off the pump's power supply and verify absence of voltage before working on the contactor or relay; never return the equipment to use with the overload protection bypassed, and functionally test the circuit before restart.
Lesson: A component named on a drawing as a 'current limiter' may simply be a standard thermal overload relay — identify the actual part and its setting before ordering, and always replace protection with an equivalent unit, never a bypass.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.