Siemens Symphony: reconstructor recovery rebuilds only the system drive
What happened
During a VA40A software recovery on the HP Z420-based reconstructor of a Siemens Symphony MRI, only the system drive C was rebuilt when two hard drives were fitted, and the expected partition layout was never created.
What was checked
- The engineers ran the installation and recovery with different numbers of hard drives and compared the results.
- Only the configuration with three drives produced the full partition layout and file set reported in the case.
Cause
In this case the VA40A installer expected three physical drives in the workstation: only with three drives present did it create the intended partitions and copy the required files, while with two it stopped after the system drive. The third drive remaining unpartitioned after installation is normal for this setup.
What was done
The engineers fitted three hard drives in the Z420 before starting the recovery; with that configuration the VA40A installation completed with the correct partition layout.
Safety
The MR magnet field is always on: follow the site's MR safety rules for tools and loose metal near the magnet, and complete the manufacturer's self-test and image checks before returning the scanner to clinical use.
Lesson: Before recovering OEM software, confirm the hardware prerequisites the installer expects — including drive count and layout — otherwise it can quietly do only part of the job.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.