Siemens Symphony: scan start errors on the host
What happened
When a scan was started, errors appeared on the host computer. No specific error codes were reported in the case.
What was checked
- The engineers first asked whether a disk had recently been added to the system or the system image had been re-imaged, since such changes can shift drive letters.
- They then compared the drive letter of the rawdata disk on the reconstructor with the path the reconstruction software expects.
Cause
The drive letter of the rawdata disk on the reconstructor had changed on its own, so at scan start the software could no longer find the raw data path.
What was done
The engineers restored the correct drive letter for the rawdata disk on the reconstructor, after which scans started normally. Noted in the case: a failing disk can produce the same picture, so disk health is worth checking before closing the job.
Safety
Before opening the host or reconstructor to inspect or swap disks, power the unit down and follow normal electrical isolation practice, and return the system to scanning only after it passes its usual functional checks.
Lesson: When a scanner throws errors at scan start, first confirm that drive letters and data paths on the reconstruction side still match the configuration before hunting for deeper faults.