Academy / Field fault notes / Siemens Symphony
Siemens · Symphony

Siemens Symphony: scan start errors on the host

What happened

When a scan was started, errors appeared on the host computer. No specific error codes were reported in the case.

What was checked

Cause

The drive letter of the rawdata disk on the reconstructor had changed on its own, so at scan start the software could no longer find the raw data path.

What was done

The engineers restored the correct drive letter for the rawdata disk on the reconstructor, after which scans started normally. Noted in the case: a failing disk can produce the same picture, so disk health is worth checking before closing the job.

Safety

Before opening the host or reconstructor to inspect or swap disks, power the unit down and follow normal electrical isolation practice, and return the system to scanning only after it passes its usual functional checks.

Lesson: When a scanner throws errors at scan start, first confirm that drive letters and data paths on the reconstruction side still match the configuration before hunting for deeper faults.