Siemens · Symphony

Siemens Symphony: signal shading at image edges on thoracic spine localizers

What happened

On localizers for thoracic spine scans, the image darkened (signal drop-off) at the top and bottom edges. Distortion over the cervical region made it impossible to plan the main sequences; when the patient was centered higher, the darkening remained only at the bottom.

What was checked

Cause

The engineers found that an oversized FOV pushed the top and bottom of the image into the low-homogeneity edge zone of the magnet, and incorrect patient centering placed the cervical spine right in that roll-off. As an additional finding, reported in the case, the system turned out to carry a gradient/bore assembly from a Siemens Sonata with a shorter usable volume.

What was done

Center the patient correctly — on Siemens systems, roughly a palm below the jugular notch per the case — fine-tune the offset in the protocol if needed, keep the field short and bring the FOV down to the working range for thoracic imaging. No parts or hardware work were involved; the outcome was reported as solved.

Safety

The static field of the main magnet is always on — keep ferromagnetic tools and equipment out of the scan room and follow the site's MRI safety procedures for any work at the bore.

Lesson: Edge shading on localizers is usually a positioning and coverage problem — check patient centering and FOV size before suspecting hardware.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.