Siemens Symphony: unexplained coolant loss in the secondary water circuit

What happened

The secondary cooling circuit kept needing top-up — roughly a litre of distilled water a day as reported in the case — while no visible leaks could be found in the cabinets or under the magnet covers.

What was checked

Cause

A connection inside the filter box was dripping into the partition between the equipment room and the magnet enclosure, so the water never showed where people normally look. The engineers also suspected a crack in the RCA heat exchanger between the circuits, a known source of this kind of slow loss.

What was done

Dry paper sheets laid under the filter box connections picked up wet spots and pointed straight to the dripping joint, which was then fixed; in this case the RCA heat exchanger was also replaced as a suspected additional source, and the engineers recommend checking or replacing it whenever such slow loss appears.

Safety

Shut down and depressurise the cooling circuit before opening any connection, and make sure the area around the repaired joint is completely dry and electrical safety checks are passed before returning the system to use.

Lesson: When distilled water disappears with no visible leak, look for hidden paths — a slow drip inside the filter box or a crack between circuits in the RCA heat exchanger — and dry paper under the joints will show what the eye cannot.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.