Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Siemens Syngo MR (высокопольный МРТ)
Siemens · Syngo MR (высокопольный МРТ)

Siemens Syngo MR (high-field MRI): worn workstation SAS drives, planned SSD swap

What happened

The original 147 GB SAS drives (MBA3147RC, as reported in the case) in the SyngoMR workstation were worn out, and the engineers planned to replace them with SSDs. The open question was whether migrating the existing Windows 7 image from the old disks would go through without errors.

What was checked

Cause

The stock SAS drives had simply aged and worn out; no other fault was reported in the case.

What was done

The worn SAS drives were replaced with SSDs and the existing Windows 7 system image was transferred onto them. In this case a colleague reported the same swap done on both a workstation and an image reconstructor, running without problems for almost two years as reported. Treat this as one reported case, not a guaranteed recipe.

Safety

Power the workstation down and isolate it from mains before opening it, use ESD precautions when handling drives, and remember the MR magnet field is always on — keep ferromagnetic tools out of the magnet room, and confirm full system checks before returning the scanner to clinical use.

Lesson: An ageing medical workstation can often be revived by swapping worn disks for modern SSDs and migrating the original system image, but such a swap is a case-by-case decision that must be proven with full checks before clinical use.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes