Siemens UMAS: blue screen during checkup

What happened

Mid-way through a system checkup the screen went to a blue screen of death; after a reboot the checkup would not pass, and the error messages gave no useful direction.

What was checked

Cause

The 5 V supply sagged because of losses across oxidised and loose contacts between the lambda block and the UMAS; the voltage arriving at the board was too low for reliable operation.

What was done

The engineers re-seated and cleaned the contacts with contact cleaner and reconnected the path; the 5 V LED then lit as expected. The error left behind after the blue screen at the Defective Channel stage was cleared with the channel-tuning procedure (Tune Defective Channel) — a procedure from the manufacturer's service documentation, done by authorised personnel. After this the checkup passed and the fault was resolved. Case note: the UMAS2/UMAS Plus has no fan by design, and its board is reportedly often damaged when someone retrofits one.

Safety

De-energise the system and let the supplies discharge before opening or re-terminating power harnesses, and repeat the manufacturer's electrical safety tests (protective earth continuity and leakage currents) before returning the system to clinical use.

Lesson: When error messages are uninformative, measure supply voltages at the point of use — a rail can look fine at the source and still arrive low after oxidised or loose contacts.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.