Siesta Whispa: leak alarm from a loose bellows rim

What happened

The machine flagged a leak in the breathing bellows. On inspection, the upper rim of the bellows sat loosely on its seat, so the assembly would not hold a seal.

What was checked

Cause

The upper rim of the bellows had stretched and lost its tight press-fit on the seat, so the circuit leaked at that joint. The engineers treated the stretched rim as the probable cause of the alarm in this case, not as a general rule for the model.

The cause is probable and was not fully confirmed.

What was done

Stopgap reported in the case: a couple of ordinary stationery rubber bands were slipped over the rim to cinch it down against the seat — purely a temporary clamp. The proper repair was a new bellows from the preventive-maintenance kit; reshaping the old rim was discussed and rejected, and the case recommended replacing the bellows once a year as part of routine maintenance. The outcome was reported as probable, not fully confirmed.

Safety

Before the machine returns to clinical use, run a full breathing-system leak test and confirm correct delivered volumes, and treat any improvised clamping as strictly temporary — non-approved materials in the gas path can fail in an oxygen-rich atmosphere.

Lesson: Before condemning a bellows for a leak, check whether the rim still grips its seat — and treat any improvised fix as a stopgap only until a new bellows is fitted per the maintenance schedule.