SMP SMP-21/01-R-D: volume error during calibration
What happened
The spirometer kept reporting a volume error during calibration, and the engineer asked whether the inspiration/expiration coefficients could be corrected directly.
What was checked
- Ran through the calibration procedure again to rule out a technique error
- Inspected the calibration syringe and found it jamming, with the grease drying out and rolling off the walls
- Tried calibration with a partial syringe stroke, increasing the swept volume step by step until a normal full calibration passed
Cause
There is no direct way to edit the inspiration/expiration coefficients; the coefficient drifts after failed calibrations, and the underlying problem in this case was a dry, jamming calibration syringe delivering inconsistent volumes — the engineers found this in one case, not as a rule.
The cause is probable and was not fully confirmed.
What was done
Service the calibration syringe — clean and re-lubricate it per the manufacturer's documentation — and when the reported volume reads high, calibrate with a partial stroke and gradually increase the volume until calibration passes normally. The outcome was reported as probable, not confirmed on the spot.
Safety
Before returning the spirometer to clinical use, disinfect the breathing circuit and all patient-contacting parts and confirm the device passes an electrical safety test.
Lesson: When calibration fails with a volume error, service and verify the calibration syringe itself before suspecting the analyser or looking for coefficient corrections.
Covered in the course: Fault-Finding Method for Medical Equipment →