Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / СМП СМП-21/01-Р-Д
СМП · СМП-21/01-Р-Д

SMP SMP-21/01-R-D: volume error during calibration

What happened

The spirometer kept reporting a volume error during calibration, and the engineer asked whether the inspiration/expiration coefficients could be corrected directly.

What was checked

Cause

There is no direct way to edit the inspiration/expiration coefficients; the coefficient drifts after failed calibrations, and the underlying problem in this case was a dry, jamming calibration syringe delivering inconsistent volumes — the engineers found this in one case, not as a rule.

The cause is probable and was not fully confirmed.

What was done

Service the calibration syringe — clean and re-lubricate it per the manufacturer's documentation — and when the reported volume reads high, calibrate with a partial stroke and gradually increase the volume until calibration passes normally. The outcome was reported as probable, not confirmed on the spot.

Safety

Before returning the spirometer to clinical use, disinfect the breathing circuit and all patient-contacting parts and confirm the device passes an electrical safety test.

Lesson: When calibration fails with a volume error, service and verify the calibration syringe itself before suspecting the analyser or looking for coefficient corrections.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes