SonoScape S40: built-in printer stopped printing after menu changes
What happened
The built-in printer of the system stopped printing after settings in the system menu had been changed. No error codes were reported in the case.
What was checked
- Check the printer settings in the system menu; the engineers found that a different printer model was selected than the one actually installed.
- Check the service documentation for the driver installation procedure; in this case it was not required once the correct model was set.
Cause
A wrong printer model had been selected in the settings, so the system was no longer configured for the built-in printer and output stopped.
What was done
The correct printer model was restored in the settings, and printing resumed. No parts were replaced and no driver installation was needed.
Safety
Before opening the printer compartment, power the system down and disconnect it from the mains, and confirm normal printing before returning the unit to clinical use.
Lesson: When a built-in device stops working right after menu changes, verify the configuration in the settings before suspecting hardware.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.