Sonoscape S6: intermittent no-boot on power-up
What happened
On power-up the button backlight comes on and the processor board has supply voltage, but boot stalls before the loader stage. The fault is intermittent: as reported in the case, the machine starts normally only about one attempt in twenty.
What was checked
- The engineers confirmed supply voltage at the processor and tested the RAM — no errors found.
- With power and memory clean, the remaining suspects for a pre-loader hang are the CPU fan (check that it spins up and reaches rated speed) and the hard drive (a spinning HDD, not an SSD): read the drive's health data and listen for retries or clicks.
Cause
Not confirmed in this case. On similar machines with the same picture, the cause turned out to be a CPU fan that failed to reach operating speed or a dying hard drive — either can stall boot before the loader, and the intermittent pattern fits both.
The cause is probable and was not fully confirmed.
What was done
No confirmed repair in this case; the outcome is probable. Practical path: verify the CPU fan and replace it if it does not reach speed, then check the drive and replace the worn HDD; any system software reload afterwards is a procedure from the manufacturer's service documentation, done by authorised personnel.
Safety
Disconnect the unit from mains and allow internal capacitors to discharge before opening it, and complete the electrical safety test before returning the machine to clinical use.
Lesson: An intermittent hang before the loader with clean power and RAM points first at the fan and the boot drive, so test them before suspecting the processor board.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.