Sonoscape S8-8000: hangs during boot

What happened

The scanner freezes partway through startup and never reaches the working screen; a similar hang was reported on an S20 Pro.

What was checked

Cause

Not confirmed in the case. The engineers suspected corruption of the ReiserFS file system on the system partition of the internal drive, stalling the boot process.

The cause is probable and was not fully confirmed.

What was done

Probable fix, not yet confirmed: with a drive image taken first, run reiserfsck --check on the system partition and, if errors are reported, reiserfsck -S --rebuild-tree to repair the file system; the alternative is a clean software reinstall from the manufacturer's installation USB, issued against the unit's serial number. The engineers had not yet confirmed which route resolved the fault.

Safety

Isolate the unit from the mains before opening it to access the internal drive, and complete the manufacturer's electrical safety test before returning it to clinical use.

Lesson: When a boot hang points to file system damage, image the drive before touching it, then repair the file system or reinstall software only through the manufacturer's official channel before blaming hardware.