Sonosite G40: shuts down with short-circuit fault seconds after power-on

What happened

The ultrasound ran for about 10 seconds (as reported in the case) and then shut down with a short-circuit fault.

What was checked

Cause

The engineers found a shorted (broken-down) small ceramic capacitor in the 12 V circuit of the built-in DC-DC module; it pulled the rail low and triggered the short-circuit protection, which shut the system down.

What was done

Replacing the damaged ceramic capacitor on the DC-DC module restored the 12 V rail; the fault cleared and, as reported in the case, the scanner was returned to service.

Safety

Unplug the unit and let the power supply capacitors discharge before touching the power modules, and run the electrical safety test before returning the scanner to clinical use.

Lesson: When an ultrasound dies with a short-circuit fault seconds after power-on, measure the DC-DC rails first — a single shorted ceramic capacitor can be the whole fault.