Sony · DS5302

Sony DS5302: printer keeps alerting on an empty upper feed tray

What happened

The upper feed tray had to be taken out of use and left empty, but the printer kept complaining about the empty tray on every start.

What was checked

Cause

Not a fault of the printer: the model has no supported way to leave the upper tray disabled, so the printer treats an empty tray as an alert condition on every start.

What was done

The engineers disconnected the RF-tag reader connector of the upper tray (accessible from below). At the next power-up the printer itself reported the missing reader and marked that tray as disabled, and the empty-tray alerts stopped. Solved in this case; since it is not a documented setting, repeat such work only per the manufacturer's service documentation and by authorised personnel.

Safety

Unplug the printer from the mains before reaching for internal connectors, and after such work print a test image and check its quality before returning the printer to clinical use.

Lesson: It is cleaner to let the device itself mark an unused option as unavailable than to silence its alerts with leftover media.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.