Sony · UP-DF550

Sony UP-DF550: printer rejects both film trays after a film change

What happened

As reported in the case: right after a new 14x17 film pack was loaded, the printer refused the tray with 'CHECK TRAY'; soon after, the second 8x10 tray was also rejected, and both trays showed 'Check'.

What was checked

Cause

The engineers found a sheet of film jammed in the feed mechanism — it had slipped during the film change and blocked the feed path, so the printer could not accept either tray.

What was done

Both trays were removed, the jammed film was located and cleared from the feed mechanisms, and the trays were reinstalled. Reported outcome: solved — the printer accepted both trays again.

Safety

Unplug the printer from the mains before reaching into the transport: jammed film has sharp edges and the drive rollers can pinch fingers.

Lesson: When a film printer rejects tray after tray right after a film change, look for a jammed sheet in the feed mechanism before suspecting the trays or electronics.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.